Previous Page  11 / 21 Next Page
Information
Show Menu
Previous Page 11 / 21 Next Page
Page Background

Page 42

Notes:

conferenceseries

.com

Volume 9

Journal of Analytical & Bioanalytical Techniques

Mass Spectra & Analytika 2018

September 19-20, 2018

September 19-20, 2018 Singapore

World Congress on

Mass Spectrometry and Analytical Techniques

J Anal Bioanal Tech 2018, Volume 9

DOI: 10.4172/2155-9872-C2-031

Analytical characteristics of nanoelectrospray operated under super-atmospheric pressure

Md Matiur Rahman

1

and Lee Chuin Chen

2

1

Beijing Institute of Technology, China

2

University of Yamanashi, Japan

H

igh-pressure nanoelectrospray ionization (nanoESI) source is a recently developed technique in which the electrospray

ionization is generated inside an enclosed chamber with gas pressure higher than the atmospheric pressure. In this paper,

the performance of nanoESI under different gas pressures, emitter position, ion inlet temperature, additive for desalination are

presented. Under a pressure of 2 bars, the nanoESI is almost eased from the electrical discharge problem and that offers a wider

tuning window for the emitter potential to produces a higher and more stable ion signal. With optimized ion inlet temperature,

the high-pressure operation facilitates the generation of ion species of higher charge-state from the highly aqueous solution

and produced less sodium adducts. A preparation method for the high-throughput analysis of raw biological samples using

disposable plastic nanoESI emitter could be described.

matiurrahmanbot@yahoo.com