The new high sensitive method of determination of toxicants in air has been developed. The combination of solid phase
microextraction and liquid phase microextraction is in the base of the method. At first it is necessary to carry out pre
concentration of impurities from the air to crystal structure of ice, which is formed from atmospheric moisture. Thus, the water
contained in the analyzed air, acts as a collector of impurities. Condensation concentration is carried out by aspiration of air through
the tubes of cryoconcentrator. The formed ice has been thawed and water condensate, containing air pollution, has been collected. Pre
concentration of impurities from the condensate was carried out by liquid-phase micro extraction with ultrasonic nano dispersion
of extractant. This approach is based on the distribution of trace amounts of suspended submicron particles of extractant throughout
the volume of the analyzed sample. Due to the increase of the mass transfer surface in 104-105 units of magnitude, the equilibrium
in the water-extractant system was established almost immediately. Aggregation of extractant phase was carried using centrifugation.
Analysis of the extract was carried out by gas chromatography-mass spectrometry. Obtained integral coefficients concentration
of impurities from the air to the extract is 104-105. Reached the limits of detection of high toxic impurities (such as benzo (a)
pyrene) is at the level of 10-5 ?¼g m-3, which is at the level of the worldâ??s leading results. The developed method will solve many
problems of traditional methods of sample preparation of air samples. Using of a condensation method solves typical for solid-phase
extraction techniques problem of sorbent background. In this case, the adsorbent is moisture of analyzed air, i.e. a substance known
to not contain impurities. Thus, there is no need for the procedure of desorption, which is time-consuming and requires expensive
extractants. Impurities from the volume sampled air moving in a few micro liters of extractant. Due to this, 20-25% of the sample,
containing the impurities of analyzed air enters to gas chromatography-mass spectrometer. In traditional methods of solid-phase
extraction, this value is 0.01-0.001%. This fact has allowed to reduce the amount of volume sampled air, to increase the speed of
analysis and to reduce the detection limit. The developed method is consistent with the concept of â??green chemistryâ? because one
solves the problem of disposing of toxic extractants.